Terumo Advanced Perfusion System 1 Temperature Monitoring Systems, Terumo Cardiovascular Systems, Ann Arbor, MI; Catalog number 802114. For use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures.

FDA Recall Z-0672-2009 · Posted 2009-01-29

Recall StatusTerminated
Product CodeDTQ
Date Initiated2007-11-20
Date Terminated2009-08-06
Product Quantity1219 (total)

Reason for Recall

The system may malfunction due to t-filter corrosion on an internal circuit board.

Firm Action

U.S. consignees were notified of the problem by letter on 11/16/07. International affiliates were informed via email on 11/21/07. Consignees were informed that Terumo would schedule appointments to install new pressure modules and reconfigure the systems beginning in Feb. 2008 or that parts would be shipped to consignees wishing to perform the work themselves. For additional information, contact Terumo CVS Customer Service at 1-800-521-2818.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.