Terminated

iTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 - Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

FDA Recall Z-0672-2020 · Posted None

Recalling FirmConformis, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2019-10-25
Date Terminated2020-05-20
Product Quantity1 unit

Reason for Recall

Medial and lateral polyethylene inserts did not securely fit into the cobalt chrome tray implant.

Firm Action

Conformis contacted surgeon via phone on 10/16/19 with the replacement details. Replacement inserts that match the implanted tibial tray interlock geometry were shipped to the affected customer.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution in the state of GA.