Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Anti-Mitochondrial Antibody Kit

FDA Recall Z-0673-05 · Posted 2005-04-06

Recalling FirmDiasorin Inc.
Recall StatusTerminated
Product CodeDBM
Date Initiated2004-04-12
Date Terminated2006-07-22
Product Quantity8

Reason for Recall

The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed.

Firm Action

A notification letter and effectiveness check form was sent to consignees. This letter recommended that users manually verify results using the specifications outlined in the package insert until a new software version is installed.

Root Cause

Other

Distribution Pattern

CA, CT, KS, MA & WA