Prostate Biopsy Needle, Model: NAC-1825BB, Soxe 18 GA x 25 CM, Single use, Ethylene sterilized, Non-pyrogenic, Remington Medical, Inc., Alpharetta, GA 30005.
FDA Recall Z-0673-2007 · Posted 2007-03-29
Advertisement
Recalling FirmRemington Medical Inc.
Recall StatusTerminated
Product CodeKNW
Date Initiated2007-01-19
Date Terminated2007-05-11
Product Quantity192 units
Reason for Recall
Pouches were not sealed prior to sterilization.
Firm Action
Consignees were contacted by telephone on 01/19/2007. Each customer was informed the sterility of the product may be compromised as a result of unsealed pouches and requested to visually check their inventory.
Root Cause
Other
Distribution Pattern
Nationwide.