Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.

FDA Recall Z-0674-05 · Posted 2005-04-06

Recalling FirmDiasorin Inc.
Recall StatusTerminated
Product CodeMSV
Date Initiated2004-10-26
Date Terminated2006-07-22
Product Quantity7

Reason for Recall

Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to 15 U/ml.

Firm Action

Customers were sent notification letters and effectiveness check form. recommended users manually veify positve results (RSLT), using product insert specifications.

Root Cause

Other

Distribution Pattern

CT, MI, MT, UT & CANADA