Medical Air Compressor Model #8413419
FDA Recall Z-0674-2007 · Posted 2007-04-05
Advertisement
Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeBTI
Date Initiated2007-01-26
Date Terminated2007-06-15
Product Quantity67 units
Reason for Recall
Compressors not providing sufficient supply pressure to connected medical ventilator
Firm Action
The recalling firm issued a Medical Device Recall Letter dated 1/22/07 to their customers via Certified Mail. An authorized service representative will inspect each potentially affected device. If a device is found to have an affected dryer assembly, it will be repaired free of charge.
Root Cause
Other
Distribution Pattern
Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.