Terminated

Medical Air Compressor Model #8413419

FDA Recall Z-0674-2007 · Posted 2007-04-05

Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeBTI
Date Initiated2007-01-26
Date Terminated2007-06-15
Product Quantity67 units

Reason for Recall

Compressors not providing sufficient supply pressure to connected medical ventilator

Firm Action

The recalling firm issued a Medical Device Recall Letter dated 1/22/07 to their customers via Certified Mail. An authorized service representative will inspect each potentially affected device. If a device is found to have an affected dryer assembly, it will be repaired free of charge.

Root Cause

Other

Distribution Pattern

Nationwide.