Terminated

Gemcal Reference Serum (magnesium calibrator). Calibrator, Multianalyte Mixture. Level 1 and Level 2 Chemistry Control.

FDA Recall Z-0676-06 · Posted 2006-03-23

Recalling FirmAlfa Wassermann, Inc
Recall StatusTerminated
Product CodeJIX
Date Initiated2006-01-26
Date Terminated2006-04-18
Product Quantity22,987 kits

Reason for Recall

Inaccurate test results. The magnesium values for the A31/A41 Gemcal Serum calibrator gave control results that were biased approx. 20% high.

Firm Action

Field correction letters were sent to the firm's customers via FAX on 1/14/2006-1/17/2006 and via UPS on 1/17/2006-1/19/2006. An updated letter was sent out via FAX on 1/27,30,31/2006 and via UPS on 1/30/2006 and 2/1/2006.

Root Cause

Other

Distribution Pattern

The kits were sent nationwide. The product was shipped to Alfa Wasserman's sister firm in Wergen, Netherlands. There are several US Navy accounts.