Open, Classified

BrightView X, Gamma Camera, Product Code 882478

FDA Recall Z-0678-2024 · Posted 2024-01-05

Recalling FirmPhilips North America
Recall StatusOpen, Classified
Product CodeKPS
Date Initiated2023-11-30
Product Quantity59 US; 59 OUS

Reason for Recall

While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.

Firm Action

On November 30, 2023, the firm notified customers of the issue through Urgent Medical Device Correction Letters. Users were instructed to monitor the patient during system movement when Pre-Programed Motion is in use and review the warning and caution guidance information provided with the recall notice, and which is also outlined in the Instructions for Use. Philips will contact you to schedule a time for a field service engineer to visit your site and install a technical solution to resolve this issue. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).

Root Cause

Device Design

Distribution Pattern

Domestic distribution nationwide. Foreign distribution worldwide.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.