Recall StatusOpen, Classified
Date Initiated2024-11-20
Product Quantity2,365 units
Reason for Recall
A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.
Firm Action
Nova Biomedical notified consignees via letter on 11/20/2024. Consignees were instructed that all current Nova StatStrip customers that have gone Live with their new meters recently received a meter software update (v.0.0.13.45 or above) from Nova Biomedical to eliminate the potential risk described above. All remaining customers that have been shipped StatStrip meters impacted by this field correction, will have their meter software updated by Nova Biomedical or local dealer prior to going "Live" with their new meters.
Root Cause
Software design
Distribution Pattern
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.