Terminated

3-Way Standardbore Stopcock. Product Code: AMS-100.

FDA Recall Z-0682-2018 · Posted None

Recall StatusTerminated
Product CodeFMG
Date Initiated2017-07-24
Date Terminated2019-11-07
Product Quantity200 units

Reason for Recall

Vygon is conducting a voluntary recall of the products and lots listed in this recall, being initiated because our Tyvek supplier has notified Vygon of issues with their product, specifically the Tyvek may not be coated completely causing gaps in the adhesive used to seal the product packaging. This potential lack of adhesive may affect the sterility of the product.

Firm Action

Vygon sent recall letters dated July 24, 2017, to customers via email and fax with read receipts and delivery confirmations. The recall letter informed customers of the reason for recall, product description, code and lot number, and instructions to customers, with a Recall Acknowledgement and Inventory Return Form provided/accompanying the recall letter.

Root Cause

Under Investigation by firm

Distribution Pattern

U.S.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.