Terminated

Philips QCPR Meter - used with FR3 Product Usage: The Philips CPR meter is intended for use with the Philips HeartStart FR3 automated external defibrillators (FR3) with PR2.0 or higher software or on HeartStart MRx devices with software version F.01/R.01 or F.02/R.02 installed. The Q-CPR option provides feedback designed to encourage rescuers to perform resuscitation in accordance with AHA/ERC guidelines for chest compression rate, depth, and duty cycle and ventilation rate, volume and flow rate (inflation time).

FDA Recall Z-0683-2018 · Posted 2018-02-22

Recall StatusTerminated
Product CodeLIX
Date Initiated2018-01-29
Date Terminated2020-05-05
Product Quantity28094 in total

Reason for Recall

Inform users of correct placement of the disposable adhesive pad on the QCPR meter and inform users of injuries associated with proper performance of CPR. As required by Consent Decree of Permanent Injunction No. 1:17-cv-11955-DJC, entered October 31, 2017 between United States of America and Philips North America

Firm Action

Philips sent an Medical Device Recall letter dated January 19, 2018 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately return the attached reply card. For question contact your local Philips representative.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US Nationwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.