Open, Classified

Sensis Vibe Hemo

FDA Recall Z-0683-2024 · Posted 2024-01-09

Recall StatusOpen, Classified
Product CodeDQK
Date Initiated2023-11-29
Product Quantity331 units

Reason for Recall

The possibility of the Sensis documentation functionality application to crash.

Firm Action

The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 11/29/23 by mail to consignees to notify them of this recall event. In the case this failure mode occurs during use, consignees are asked to continue the exam by opening a new study for the same patient without adding any Type 1 reporting events. It is instructed that consignees can compensate this by adding free text comments (procedure notes) only and manually inserting the desired information in those comments. The recalling firm will deploy a software update via Update Instruction AX010/23/S to correct the error. A member of the service organization will contact consignees to schedule the update. Consignees can contact the service organization at 1-800-888-7436 to schedule an earlier appointment. The recall notice is to be disseminated to all users of affected products within consignee organizations and others who may need to be informed. Consignees are to return the provided response form by email to [email protected] within 14 days of receipt.

Root Cause

Software design

Distribution Pattern

Domestic: CA, CT, & FL.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.