CryoValve, Pulmonary Valve and Conduit

FDA Recall Z-0684-05 · Posted 2005-04-06

Recalling FirmCryolife Inc
Recall StatusTerminated
Product CodeMIE
Date Initiated2004-12-17
Date Terminated2005-06-01
Product Quantity1 tissue

Reason for Recall

During a retrospective review, CryoLife noted that the donor tissue exceeded the acceptable warm ischemic time when utilizing the 'Last Seen Alive' time in their calculation.

Firm Action

Consignees were notified by letters dated 12/17/2004, 01/03/2005 and 01/19/2005.

Root Cause

Other