Swiss Lithocast Ultra Ultrasound Handpiece (EMS- PN:EL-236) used in combination with Swiss Lithoclast Ultra System .
FDA Recall Z-0684-2007 · Posted 2007-03-30
Advertisement
Recalling FirmE.M.S. - Electro Medical Systems SA
Recall StatusTerminated
Product CodeFFK
Date Initiated2005-03-16
Date Terminated2016-02-02
Product Quantity160 units
Reason for Recall
Handpiece may become dislodged causing cable damage and possible electric shock.
Firm Action
EMS Lyon, France (manufacturer) notified the single US Distributor by letter to contact their accounts and return device for replacement. Boston Scientific, Natick, Ma conducted the sub-recall by letter dated 4/22/05 to all USA accounts.
Root Cause
Other
Distribution Pattern
Nationwide
Related FDA 510(k) Records