Innova Self-Expanding Stent System, 180mm. Labeled with Material Number and description as follows: Product Usage: The Innova Vascular Self-Expanding Stent System is indicated to improve luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery (PPA) with reference vessel diameters from 4.0 mm to 7.0 mm and lesion lengths up to 190 mm. a. H74939180051870 (Material Description Innova-EU 5 x 180 x 75); b. H74939180061870 (Material Description Innova-EU 6 x 180 x 75); c. H74939180071870 (Material Description Innova-EU 7 x 180 x 75); d. H74939180071870 (Material Description Innova-EU 8 x 180 x 75); e. H74939181051830 (Material Description Innova-EU 5 x 180 x 130); f. H74939181061830 (Material Description Innova-EU 6 x 180 x 130); g. H74939181071830 (Material Description Innova-EU 7 x 180 x 130); h. H74939181081830 (Material Description Innova-EU 8 x 180 x 130); i. H74939259051830 (Material Description Innova-China 5 x 180 x 130); j. H74939259061830 (Material Description Innova-China 6 x 180 x 130); k. H74939259071830 (Material Description Innova-China 7 x 180 x 130); l. H74939259081830 (Material Description Innova-China 8 x 180 x 130); m. H74939259081870 (Material Description Innova-China 8 x 180 x 75); n.H74939293051830 (Material Description Innova 5 x 180 x 130); o. H74939293051870 (Material Description Innova 5 x 180 x 75); p. H74939293061830 (Material Description Innova 6 x 180 x 130); q. H74939293061870 (Material Description Innova 6 x 180 x 75); r. H74939293071830 (Material Description Innova 7 x 180 x 130); s. H74939293071870 (Material Description Innova 7 x 180 x 75); t. H74939293081830 (Material Description Innova 8 x 180 x 130); u. H74939293081870 (Material Description Innova 8 x 180 x 75);

FDA Recall Z-0686-2018 · Posted 2018-02-22

Recall StatusOpen, Classified
Product CodeNIP
Date Initiated2017-11-06
Product Quantity22,916 units total

Reason for Recall

Stent possibly unable to be fully released from the delivery system.

Firm Action

Boston Scientific sent an Urgent Medical Device letter dated November 6, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers\ were instructed to immediately discontinue use of the product and remove all affected products from inventory, complete and return the Verification Form, return affected products and provide the notification to any healthcare professionals affected.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - US Nationwide Distributed domestically to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV. Distributed internationally to Canada

Official Source

View FDA Recall →