Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216
FDA Recall Z-0686-2025 · Posted 2024-12-17
Advertisement
Recalling FirmPhilips North America Llc
Recall StatusOpen, Classified
Product CodeMHX
Date Initiated2024-11-20
Product Quantity4,811 units
Reason for Recall
The following four issues are included:
1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network.
2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link.
3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts.
4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.
Firm Action
Consignees were notified on about 11/20/2024 via certified mail. Consignees were instructed to identify affected units, share the communication with clinical staff, and place the Urgent Medical Device Correction with the documentation of the Philips Smart-hopping 2.0 AP 1.4 GHz and associated devices. A Philips representative will contact customers to arrange a software upgrade to E.19.02 (5.0.19.02) or greater to correct the issues. Distributors were instructed to contact customers and provide the Medical Device Correction to each.
Root Cause
Under Investigation by firm
Distribution Pattern
Nationwide distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.