HTA Console System, 230 Volt , French, Product Number: 56001FO (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)

FDA Recall Z-0690-2007 · Posted 2007-03-31

Recall StatusTerminated
Product CodeMNB
Date Initiated2007-01-24
Date Terminated2007-12-12
Product Quantity2 units

Reason for Recall

Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment

Firm Action

Boston Scientific notiifed accounts by letter dated January 24, 2007 via federal Express Mail. Accounts were requested to examine inventory, remove and return product.

Root Cause

Other

Distribution Pattern

Worldwide-USA, UK, Denmark, Italy, and France