Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Biofinity XR Toric contact lenses 8.7/14.5 - 6PK Biofinity XR Toric (comfilcon A) Soft (Hydrophilic) Contact Lenses are indicated for the correction of ametropia (myopia or hyperopia) with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -5.75 diopters. The Biofinity (comfilcon A) Soft (Hydrophilic) Contact Lenses have been approved for extended wear for up to 6 nights /7 days of continuous wear. It is recommended that the contact lens wearer first be evaluated on a daily wear schedule. If successful, then a gradual introduction of extended wear can be followed as determined by the prescribing Eye Care Practitioner.

FDA Recall Z-0691-2018 · Posted None

Recalling FirmCooperVision Inc.
Recall StatusTerminated
Product CodeLPM
Date Initiated2017-09-01
Date Terminated2018-07-30
Product Quantity120

Reason for Recall

The firm identified a labeling error affecting Made-To-Order (MTO) Biofinity XR Toric blister labels. The primary blister reads Biofinity Multifocal Toric in error, and should read Biofinity XR Toric. The outer carton correctly reads Biofinity XR Toric.

Firm Action

Notifications were sent with instructions to examine inventory, discard any affected inventory, and contact the firm for credit or replacement. For further questions, please call (585) 756-9273.

Root Cause

Employee error

Distribution Pattern

USA Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑