Ascension Silicone PIP-Silicone PIP Sz 1, Model Number: SPIP5201WW - Product Usage: The Ascension Silicone PIP implant is intended for cementless replacement of the proximal interphalangeal joint in patients with advanced osteoarthritis, post-traumatic arthritis and rheumatoid arthritis.

FDA Recall Z-0691-2020 · Posted None

Recall StatusTerminated
Product CodeKYJ
Date Initiated2019-11-04
Date Terminated2020-05-05
Product Quantity9 units

Reason for Recall

Silicone PIP Implant mislabeled as a size 1 implant on the outer packaging when in fact the package contained a size 0 implant

Firm Action

Integra notified by letter consignee via courier service, facsimile or email between November 4th and 6th, 2019, advising reason for recall, health risk and action to take: confirm that the device was actually implanted as part of the above-referenced surgery and, if so, please confirm the date of that surgery. 2. Complete the attached Acknowledgement Form and return the completed form by email or fax indicating receipt and review of this notification. Keep a copy of the form for your records. Should you have any questions regarding these instructions, please contact Customer Service Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: [email protected] Outside USA: 781-565-1401

Root Cause

Labeling mix-ups

Distribution Pattern

US Nationwide distribution in the state of CA.