Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111
FDA Recall Z-0693-2016 · Posted 2016-01-27
Advertisement
Recalling FirmFresenius Medical Care Renal Therapies Group, LLC
Recall StatusTerminated
Product CodeFKX
Date Initiated2015-12-15
Date Terminated2019-10-07
Product Quantity14,094 units
Reason for Recall
The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitoneal Volume (IIPV) (overfill).
Firm Action
Fresenius Medical Care issued on December 15, 2015, a n Important Product Notification to all affected customers informing them of the problem and how to recognize the proper latch closing. Questions regarding the contents of the Customer Notification or questions regarding the issue, contact Technical Services at 1-800-227-2572 and reference the Field Action Number FA-2015-10-W.
A reply form is included as part of the customer communication with
instructions to fill out and return via fax or scanned/emailed copy.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.