Galaxy System (GAL-001)
FDA Recall Z-0693-2025 · Posted 2024-12-18
Advertisement
Recalling FirmNoah Medical
Recall StatusOpen, Classified
Product CodeEOQ
Date Initiated2024-10-25
Product Quantity26
Reason for Recall
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
Firm Action
On 10/25/24, the firm started hand delivering correction notices to customers informing them of the following:
If your device is affected the firm will contact you to inspect and if necessary to replace the power supply.
Firm contact: [email protected], 1-888-325-6624
Root Cause
Device Design
Distribution Pattern
US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.