Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay, REF: PRD-07400
FDA Recall Z-0694-2025 · Posted 2024-12-20
Advertisement
Recalling FirmHologic, Inc.
Recall StatusOpen, Classified
Product CodeQOF
Date Initiated2024-11-08
Product Quantity3
Reason for Recall
SARS-CoV-2/Flu A/B/RSV Assay Kit may erroneously produce SARS-CoV-2 false positive results.
Firm Action
On 11/8/24, recall notices were mailed to customers asking them to do the following:
1) Discontinue use, and discard affected kits.
2) Complete and return the customer response firm via email to [email protected]
If you require technical support, contact the firm's Technical Support at 1-888-484-4747, or 1-858-410-8511, or [email protected]
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
US: AK
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.