Terminated

Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT111

FDA Recall Z-0695-06 · Posted 2006-03-29

Recall StatusTerminated
Product CodeBTT
Date Initiated2005-12-20
Date Terminated2012-04-19
Product Quantity260 boxes (2,600 pcs)

Reason for Recall

Risk of fire. The manufacturing process may have the potential to damage the insulation of the heater wire in circuits within the specified lot range. This may lead to a potential risk of fire.

Firm Action

On December 20th, 2005 the firm sent the notification letter instructing consignees to check and destroy any stock in the affected lot range. A fax-back Device Recall Response Form was included with the letter and will be used for the company''s effectiveness checks.

Root Cause

Other

Distribution Pattern

Nationwide in AR, CA, FL, IL, LA, MN, NC, OH, SD, TX, UT, & VA. International distribution