Terminated

Gambro Prismaflex Continuous Renal Replacement System, Catalog Number 602314700, Gambro Lundia AB, Sweden.

FDA Recall Z-0695-2007 · Posted 2007-03-31

Recall StatusTerminated
Product CodeKDI
Date Initiated2007-01-04
Date Terminated2012-01-03
Product Quantity153 units (Domestic)

Reason for Recall

Under certain conditions, * an excessive amount of anticoagulant may be infused into the patient. (* interruption of a self-test by an alarm and obstruction of the access lines)

Firm Action

Customers were notified by letter on 12/19/2006. They were told that Gambro would schedule the required upgrades to their equipment and not to use the current syringe for delivering anticoagulant.

Root Cause

Device Design

Distribution Pattern

Nationwide distribution -- including states of AL, CA, FL, GA, IN, MD, ME, MI, MS, NC, NH, NY, OH, PA, SC, TX, VA, and WI.