Terminated

Proteus 235 Product Usage: The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.

FDA Recall Z-0698-2019 · Posted None

Recall StatusTerminated
Product CodeLHN
Date Initiated2018-12-12
Date Terminated2020-04-15
Product Quantity7

Reason for Recall

There is a risk of detachment of the Compact Gantry Rolling Floor.

Firm Action

On December 12, 2018, the firm notified its customers of the recall via an Urgent Medical Device Correction letter. Customers were informed of the product issue. Customers were reminded that nobody should stand on the rolling floor when the gantry is rotating, as specified in the Clinical User Guide. The recalling firm stated they would take the following actions: Short-term action: IBA will replace the rod end bearings attaching the cGRF to the nozzle via a preventive maintenance on all ProteusONE sites. Long-term action: A new design of the anchoring system of the cGRF to the nozzle will be deployed on all impacted sites via an upgrade kit by end of October 2019. Customers with questions should contact Sylviane BERGER [email protected] +32 10 203 787, or Helpdesk +32 2 507 20 81 (available 24/7)

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of LA and MI. Foreign distribution to France, Japan, and UK.