Recall StatusOpen, Classified
Date Initiated2023-11-28
Product Quantity4,235 units
Reason for Recall
Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
Firm Action
On December 6, 2023, Medtronic Navigation, Inc. issued a Urgent: Medical Device Recall Notification via: UPS . Medtronic ask consignees to take the following actions:
1. Immediately locate and quarantine all unused impacted product(s). Refer to the affected lot numbers identified in Table 1 below.
2. Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form
3. Complete the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information, and email the completed form to Medtronic at
[email protected].
If the affected devices have already been utilized and/or discarded, we still ask that you complete and return the Customer Confirmation Form detailing that information.
4. This notice should be distributed to all others in your organization who should be aware, or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AK, AL, AZ, CA and the countries of Armenia, Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Luxembourg, Netherlands, New Zealand, Oman, Portugal, Spain, Switzerland, United Kingdom.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.