Tachos DR - Implantable Cardioverter Defibrillator.
FDA Recall Z-0700-03 · Posted 2003-04-01
Advertisement
Recalling FirmBiotronik Inc
Recall StatusTerminated
Product CodeLWS
Date Initiated2003-02-21
Date Terminated2003-09-08
Product Quantity240 (225 still implanted)
Reason for Recall
Potential for battery to delivery less than full amount of energy.
Firm Action
The firm issued an 'URGENT - VOLUNTARY MEDICAL DEVICE' recall letter dated 2/21/03 to the physicians. This letter explains the issue and recommends scheduling an immediate follow-up with the patients.
The letter describes conditions under which the patient should be monitored on a weekly basis (versus every 3 months), and under which conditions the device should be explanted.
Root Cause
Other
Distribution Pattern
The firm distributed devices to doctors located throughout the U.S. There is no international distribution from this firm.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.