Presource Kit Total Joint Pack containing a double-wrapped basin component. Product Usage: This kit is used on patients undergoing joint replacement knee or hip.
FDA Recall Z-0702-2019 · Posted None
Advertisement
Recalling FirmCardinal Health Inc.
Recall StatusTerminated
Product CodeLRO
Date Initiated2018-11-16
Date Terminated2019-11-04
Product Quantity182 packs
Reason for Recall
The double wrapping of this custom kit basin set is not supported by the current sterility assurance validation.
Firm Action
As only one customer is potentially impacted, in the interest of prompt action and notification, a teleconference was held on Friday, November 16, 2018 with the customer to communicate the impending actions.
During this teleconference, the customer was informed that an HHE and field corrective action was in-process and that a study was being initiated to determine that their requested packaging configuration was acceptable. The customer also was informed that they should not continue to use the impacted double wrapped basin, and the components contained within, until the study was completed.
Cardinal Health is following up on this teleconference with a written notification.
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
US state of Florida