DiaSorin GammaCoat Androstenedione - 125 I, RIA Kit for the quantitative determination of androstenedione levels in serum. Catalog No. CA-1725. Lot # 04164. DiaSorin Stillwater, Minnesota 55082-0285, U.S.A.

FDA Recall Z-0704-05 · Posted 2005-04-13

Recalling FirmDiasorin Inc.
Recall StatusTerminated
Product CodeCIZ
Date Initiated2004-05-25
Date Terminated2006-07-22
Product Quantity70 kits

Reason for Recall

Standards contained in the GammaCoat Androstenedione RIA Kit appeared yellowish in color and were gelatinous upon reconstitution. The assay validation criteria failed when the gelatinous standards were used.

Firm Action

Customers were sent notification letters informing them of the product and problem and recommended the destruction of any remaining inventory. Letter also requested the completion and return of the lower portion of the letter.

Root Cause

Other

Distribution Pattern

Massachusetts, Saudi Arabia, Sweden, Germany, Spain, Belgium, Italy, Camada. Australia, Portugal, Hungary, Israel,