Terminated

QC Sets and Panels: KWIK-STIK Cepheid Smart GBS QC Set (Live Culture). Sets include: 1. 0864E4 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 2. 0864E5 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 3. 0864E7 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 4. Negative Control 0243 Lactobacillus acidophilus derived from ATCC 4356

FDA Recall Z-0704-2022 · Posted None

Recalling FirmMicrobiologics Inc
Recall StatusTerminated
Product CodeJTR
Date Initiated2015-12-01
Date Terminated2022-07-07
Product Quantity8 units

Reason for Recall

The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in some cases. This causes a failure when used as a quality control for the Cepheid Xpert GBS and Cepheid Smart GBS systems. The failure of the quality control test causes a disruption for the testing laboratory which may result in repeated testing or the need for alternative test methods. This recall is applicable to the 8164 and 8165 sets only.

Firm Action

Microbiologics notified customers on about 12/01/2015, via letter titled "MEDICAL DEVICE RECALL." Instructions to customers included verify lots in inventory, cease use of affected product, quarantine affected product, arrange for the return of affected product, and complete and return the response form.

Root Cause

Material/Component Contamination

Distribution Pattern

US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, MN, NC, NM, NY, OH, OR, PA, TN, TX, VA, VT, WA, WI, WV, WY.