Terminated

Gambro Prisma M100 preset, Part No. 8353486.

FDA Recall Z-0705-06 · Posted 2006-04-05

Recall StatusTerminated
Product CodeKDI
Date Initiated2005-10-24
Date Terminated2012-04-17
Product Quantity28277 units

Reason for Recall

After dialysis machine is disconnected from the patient, blood leakage may occur while unloading the tubing sets from the machine.

Firm Action

Recall information was forwarded to all Gambro Sale Organizations. Notification was sent to all US consignees on November 2nd and November 3rd, 2005.

Root Cause

Other

Distribution Pattern

Nationwide, including VA facilities in AZ, CA, DC, FL, GA, MI, MO, NJ, NY, OH, OR, PA, PR, TX, UT, VA, WI. Military distribution to CA and TX. Foreign distribution to Australia, Canada, Denmark, France, Germany, Hungary, UK.