MEVION S250i

FDA Recall Z-0705-2019 · Posted None

Recall StatusTerminated
Product CodeLHN
Date Initiated2018-10-01
Date Terminated2020-04-10
Product Quantity1 unit

Reason for Recall

Treatment is allowed to continue (via partials) in situations where dDose1 interlock has tripped

Firm Action

On-site Mevion Field Service Engineer communicated with the firm and is in daily communication with the single affected consignee.

Root Cause

Software Manufacturing/Software Deployment

Distribution Pattern

DC