Completed

Elekta Unity

FDA Recall Z-0706-2019 · Posted None

Recalling FirmElekta Limited
Recall StatusCompleted
Product CodeIYE
Date Initiated2019-01-04
Product Quantity2 units

Reason for Recall

In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.

Firm Action

The firm notified their consignees about the issue on 01/04/2019 and will install additional fixation.

Root Cause

Under Investigation by firm

Distribution Pattern

TX and WI, Germany, Denmark, Netherlands, United Kingdom