Terminated

DePuy Spine VIPER POLY SCREW 5.5 8 X100MM Tl Product Code: 1867-15-899 Spinal Implant Component.

FDA Recall Z-0707-2013 · Posted 2013-01-24

Recalling FirmDePuy Spine, Inc.
Recall StatusTerminated
Product CodeNKB
Date Initiated2012-11-07
Date Terminated2013-12-24

Reason for Recall

Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head

Firm Action

DePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to [email protected]. For questions call Linda Regan @508-977-6606.

Root Cause

Component change control

Distribution Pattern

Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.