Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392
FDA Recall Z-0708-2018 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeOWB
Date Initiated2017-06-06
Date Terminated2019-05-16
Reason for Recall
Potential for liquid entry
Firm Action
On June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail.
Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by calling 610-448-6461.
Root Cause
Device Design
Distribution Pattern
U.S. Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.