Open, Classified

76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN) Appx 2.9 ml,PUR YELLOW TRANS SET,ChemoClave,CK VLV,Spiros,PURPLE CAP, REF: 011-CH3568; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext w/2 ChemoClave, Red Ring, BCV; 297cm Ext w/ChemoClave, REF: 011-CH3600; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext, w/2 ChemoClave, Red Ring, BCV; 597 cm Ext w/ChemoClave, REF: 011-CH3601; Kit, 2 5 Gang Stopcocks w/5 MicroClave Clear, Spiros; 2 Smallbore Bifuse Ext Sets, 2 MicroClave Clear, Red Ring, Check Valve, RMLL; 2 Ext Sets, MicroClave Clear Clipped Together, RMLL, REF: 011-CH3610; Admin Set w/Clave, Spiros, REF: 011-CH3772; Admin Set w/Clave, Spiros, REF: 011-CH3773; Admin Set w/Clave, Spiros, REF: 011-CH3967; 76" (193 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, Drop-In Red Cap, REF: 011-CH4011

FDA Recall Z-0708-2021 · Posted None

Recalling FirmICU Medical, Inc.
Recall StatusOpen, Classified
Product CodeONB
Date Initiated2020-08-31
Product Quantity4790

Reason for Recall

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Firm Action

On 08/31/2020, Medical Device Recall notices were mailed to customers informing them to take the following actions: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) ICU Medical has some lots of unaffected product available today and is actively increasing the amount of available inventory. In the event specific product is unavailable, consider use of the non-spinning Spiros or the ChemoLock CSTD as alternatives. Please contact customer service for product availability. 4) Return affected product using the return label provided with this letter. Contact Stericycle at 1-888-871-7108 (MF, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. 5) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle to obtain a response form. For further inquiries, please contact: Global Complaint Management, 1-844-654-7780 or [email protected]; Customer Service, 1-866-829-9025, option 8 or, [email protected], (M-F, 8am-6pm CT) On 10/07/2020, the Medical Device Recall notice, with an expanded affected device scope, was sent to customers.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Official Source

View FDA Recall →