Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Monitoring Service Application (MSA). Software to process, analyze, display, and report cardiac events in ECG data received from compatible devices.

FDA Recall Z-0708-2025 · Posted 2024-12-23

Recall StatusOpen, Classified
Product CodeDPS
Date Initiated2024-12-18
Product Quantity~130,000 with 41,282 customers (1 software copy used)

Reason for Recall

Not all Electrocardiogram (ECG) events received July 2022-July 2024 were not properly routed and subsequently reviewed due to an analysis step being disabled with the monitoring service application software.

Firm Action

On 12/18/2024, the firm sent an Urgent Medical Device Correction Letter via certified mail to customers informing them that it has been identified a software configuration issue that resulted in a percentage of ECG events received into Monitoring Service Application (MSA) were not properly routed due to the Secondary Analysis being disabled and subsequently there ECG events were not reviewed by a Cardiology Technician for inclusion in reporting. Importantly, some of the ECG events received into MSA during this period (July 2022-July 2024) met notification criteria for escalation to the ordering practitioners but were not escalated. This may have resulted in missing information in the reports or notifications you received, which could have impacted your clinical decision-making. Philips is offering to have the prescription reprocessed to include the ECG events not initially routed to a Cardiology Technician. An Updated Communication to be sent will include Actions to be taken by customers: . Review the list of Frequently Asked Questions relating to this issue. . Log onto the Prescriber Response Site at https://prs.gobio.com with the Location Code provided in the header of this letter. . Read and acknowledge the receipt of this Urgent Medical Device Correction Letter on the website . Review the list of your patients and select whether and on which patients you wish reprocessing to be performed options available include reprocessing All, None, or Selected specific patients on the site. For questions/assistance-contact Prescriber Response Line 888-521-1684.

Root Cause

Use error

Distribution Pattern

US Nationwide distribution including PR.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑