LIFEPAK 20 defibrillator/monitor

FDA Recall Z-0711-06 · Posted 2006-04-04

Recall StatusTerminated
Product CodeLDD
Date Initiated2006-03-22
Date Terminated2011-09-06
Product Quantity2678 devices

Reason for Recall

Therapy connectors may have connector sockets that do not meet design specs - short drill depth. Interruption of the connections may cause unexpected, intermittent or complete loss of pacing and defibrillation functions.

Firm Action

On 3/22/06, the firm began sending out letters dated March 2006, and flagged URGENT MEDICAL DEVICE CORRECTION. The letters advised customers of the issue and that a service representative would visit and inspect the therapy connector sockets, and replace the therapy connector and the mating cable if short drill sockets are found.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide, and Canada, Netherlands, Germany, Australia, Guam, Hong Kong, New Zealand, New Caledonia, Argentina, Brazil, Chile, Costa Rica, Mexico, Uruguay, Venezuela. Domestic distribution includes hospitals, clinics,