Empowr Knee System Product Usage: Non-porous proximal tibial baseplate implants, utilized in either posterior cruciate ligament (PCL) retaining or PCL substituting total knee arthroplasty procedure.
FDA Recall Z-0711-2017 · Posted None
Advertisement
Recalling FirmEncore Medical, Lp
Recall StatusTerminated
Product CodeJWH
Date Initiated2016-11-15
Date Terminated2017-01-20
Product Quantity7 units
Reason for Recall
A size 4 Empowr baseplate was mislabeled as a size 6. The affected lot was laser marked as M6, marked with the lot number for the size 6 minus lot, and distributed as size 6 minus.
Firm Action
The recalling firm sent an Urgent Field Safety Notice letter dated November 15, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call 760-734-3551.
Root Cause
Mixed-up of materials/components
Distribution Pattern
US Nationwide Distribution in the states of CO, IL, IN, FL, CA, AZ
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.