Axiom Artis model Interventional fluoroscopic x-ray foot switch accessory; Model Numbers: 4787797, 4787805, and 4787813.

FDA Recall Z-0711-2018 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2017-09-28
Date Terminated2020-03-09
Product Quantity142

Reason for Recall

The possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release radiation. The hand switch provided can be used to release exposures; however, fluoroscopic examinations will not be possible.

Firm Action

Letters were sent to all consignees with instructions to contact their service provider to arrange for an appointment to update the device hardware.

Root Cause

Device Design