Terminated

ECAT Emerge PET Scanners (distributed by Siemens Medical Solutions USA, Inc.), ECAT ART PET Scanners (distributed by Siemens Medical Solutions USA, Inc.) and Sceptre PET Scanners (distributed by Hitachi Medical Systems of America)

FDA Recall Z-0712-06 · Posted 2006-04-04

Recall StatusTerminated
Product CodeKPS
Date Initiated2006-03-28
Date Terminated2009-12-22
Product Quantity57 devices

Reason for Recall

There is a potential for inaccurate, elevated SUVs (standardized uptake values). On follow-up scans, the elevated SUV could be misinterpreted, and inappropriately impact a patient's treatment plan.

Firm Action

Written notification was sent out beginning 03/28/2006 to end users to advise them of the problem. The notification also included specific work instructions to aid the end user until the patch is available and the upgrade could be performed by a service representative.

Root Cause

Other

Distribution Pattern

KS, WA, CA, OH, FL, TX, Germany, Great Britain, Canada Switzerland, Turkey, Argentina, Austria, France, & Spain