ProLite Mesh - mesh, surgical, polymeric ProLite Mesh is intended for use in soft tissue deficiencies including hernia repair, traumatic or surgical wounds and chest wall reconstruction procedures requiring reinforcement with a non-absorbable supportive material.
FDA Recall Z-0712-2018 · Posted 2018-02-23
Advertisement
Recalling FirmAtrium Medical Corporation
Recall StatusTerminated
Product CodeFTL
Date Initiated2017-11-22
Date Terminated2018-07-30
Product Quantity60 units
Reason for Recall
An incorrect ProLite Mesh configuration was packaged as product code 1010306-06, with lot number 412298. ProLite Mesh manufactured as product code 1010306-04 was inadvertently placed inside the pouch of packages labeled with product code 1010306-06 and lot number 412298.
Firm Action
Please examine inventory to determine if you have any of the affected products listed on page 1. If so, please remove affected products, quarantine them, and place in a secure location. If you have affected product, please contact Getinge Customer Service at [email protected] for a Return Authorization (RA) and shipping instructions to return any affected product. Pack the product to be returned with the appropriate return documents, and using the shipping instructions provided, arrange for pick up with the designated delivery service provider. Please complete and sign the enclosed Urgent Medical Device Field Safety Notice to acknowledge receipt. Return the completed form to the following e-mail address: [email protected]
For further questions, please call (973) 709-7654.
Root Cause
Under Investigation by firm
Distribution Pattern
Internationally to Spain only.