Open, Classified

QMS Tacrolimus Calibrators, Product code 10015573

FDA Recall Z-0712-2024 · Posted 2024-01-11

Recall StatusOpen, Classified
Product CodeJIT
Date Initiated2023-11-21
Product Quantity10 sales units US; 1 sales unit OUS

Reason for Recall

Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it unsuitable for use. Use of the affected lot may results in erroneous calibration results, causing delay in testing.

Firm Action

On 21 November 2023, all consignees who were shipped the VITROS QMS Tacrolimus Calibrators Lot 74859619 expiry date: 11-May-2024 from QuidelOrtho distribution center were notified of the issue and instructed to: -Discontinue using, render unusable, and discard your inventory of QMS Tacrolimus Calibrators, Lot 74859619, which was shipped to the laboratory from QuidelOrtho. -Post the notification by the laboratory s inventory storage area until a replacement lot of QMS Tacrolimus Calibrator has been received. -Forward the notification if the affected product was distributed outside their facility. If you have questions about this recall, please contact QuidelOrtho's Global Services Organization (formerly Ortho Care) at 1-800-421-3311.

Root Cause

Storage

Distribution Pattern

Domestic distribution to IL, IN, FL, NJ, NM, TX. International distribution to Chile.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.