Terminated

SonoCalc IMT 3.0, a windows based software used in conjunction with images from high-resolution ultrasound systems (MicroMaxx, TITAN, and SonoSite180PLUS). This software generates a report with the patients intima media thickness (IMT) based on average thickness of the carotid artery.

FDA Recall Z-0713-06 · Posted 2006-04-06

Recalling FirmSonosite, Inc.
Recall StatusTerminated
Product CodeLLZ
Date Initiated2005-10-26
Date Terminated2006-04-10
Product Quantity155 units of the SonoCalc product to about 80 customers/distributors

Reason for Recall

When SonoCalc IMT 3.0 software is used with the SonoSite 180Plus ultrasound system, the software miscalculates the intima-media-thickness, a measurement of risk for cardiovascular or cerebrovascular events. Underestimates percentage of stenosis, false negative result.

Firm Action

On October 26, 2005, the company sent their customers a User Guide Addendum that told customers how to verify that the image calibration data was correct.

Root Cause

Other

Distribution Pattern

Nationwide, Kuwait, Mexico, South Africa, Chile, Spain, Japan, Puerto Rico, Australia. A list of 155 cases in which the software was sent out lists medical facilities, physicans, and SonoSite Distributors.