Medisystems 15 Gauge x 3/5 inch Buttonhole Needle Set Product Code: BH-2035PE Product Usage: Medisystems Buttonhole Needle Sets are indicated for use as an access device for dialysis and pheresis procedures using a constant site orbuttonhole method of needle insertion.

FDA Recall Z-0714-2013 · Posted 2013-01-24

Recall StatusTerminated
Product CodeFIE
Date Initiated2012-12-20
Date Terminated2013-10-10
Product Quantity26 casesx 250 needles (6500needles)

Reason for Recall

A total of 5 (ea) single needle pouches are mislabeled as 15 Ga x1" instead of 15 Ga x 3/5".

Firm Action

Medisystems sent a notification letter dated December 20, 2012 to affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to immediately open and inspect inventory for the affected products. Results of the inspection are to be provided to Medisystems and the product is to be returned to Medisystems. For questions call (978) 332-5973.

Root Cause

Labeling mix-ups

Distribution Pattern

US Nationwide Distribution in the states of NY, PA and TN.