Draeger Infinity Acute Care System Monitoring Solution

FDA Recall Z-0714-2014 · Posted 2014-01-10

Recall StatusTerminated
Product CodeMHX
Date Initiated2013-12-10
Date Terminated2015-05-20
Product Quantity439

Reason for Recall

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Firm Action

An Urgent Medical Device Recall Letter, dated December 2013, was sent to Users. Once available Drager will update components of the IACS Monitoring Solution System.

Root Cause

Software design

Distribution Pattern

US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.