Terminated

Accu-Chek Comfort Curve glucose test strips; part numbers 2030420, 2030365, 2030373, 2030381, 3000133 and 3000141.

FDA Recall Z-0715-03 · Posted 2003-04-09

Recall StatusTerminated
Product CodeLFR
Date Initiated2003-02-12
Date Terminated2003-09-15

Reason for Recall

Crack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings.

Firm Action

Urgent Product Correction letters dated 2/12/03 were issued to pharmacists, specific health care providers, and the laboratory manager at customer accounts. Letter recipients were asked to advise their patients to inspect their vial for a crack prior to each use.

Root Cause

Other

Distribution Pattern

United States, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, Guatemala, Honduras, Hong Kong, Indonesia, Japan, Malaysia, Mexico, New Zealand, Norway, Pakistan, Peru, Philippines, Portugal, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, United Kingdom and Vietnam.