Polaris Spectra System Control Unit, a component of the StealthStation S7 System. Medtronic Navigation, Louisville, CO 80027. Intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.

FDA Recall Z-0715-2011 · Posted 2010-12-17

Recall StatusTerminated
Product CodeHAW
Date Initiated2010-11-05
Date Terminated2011-05-03
Product Quantity48 devices

Reason for Recall

Electronic surgical navigation aid may lose tracking ability or stop working.

Firm Action

Customers were notified by a Product Correction Notification letter on 11/05/2010. The letter advised customers of the affected product, the potential problem, and the description of the correction. Customers were told that a Medtronic rep will visit them to install a replacement control unit. Further information is available at 800-595-9709 or 720-890-3200.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution -- US, including states of AL, AZ, CO, FL, GA, KY, MD, MN, NE, PA, TX, WI, and MD and countries of Australia, Belgium, Canada, Czech Republic, Italy, Netherlands, Poland, Portugal, Saudi Arabia, Singapore, South Korea, Spain, Turkey, and UAE.