GE Healthcare, Mobile Fluoroscopy X-Ray System, OEC Model 7600, GE Healthcare, Surgery, Salt Lake City, Utah.

FDA Recall Z-0716-2008 · Posted 2008-03-15

Recall StatusTerminated
Product CodeJAA
Date Initiated2007-09-28
Date Terminated2011-11-01
Product Quantity157 units (Domestic), 142 units (Foreign)

Reason for Recall

Unnecessary operator exposure; X-ray machine may allow unwanted x-ray exposure to the operator.

Firm Action

On or about 9/6/2007, by Customer Notification Letter, users were notified that all affected units will be corrected by GE personnel with the addition of an additional shield.

Root Cause

Other

Distribution Pattern

Worldwide- USA including states of CO, DE, FL, NJ, MO, NY, PA, and TX (with VA and Military facilities) and countries of Australia, Austria, Belgium, Brazil, Chile, China, The Czech Republic, Denmark, France, Germany, Hungary, Israel, Italy, Norway, Spain, Sweden, Switzerland, Thailand, and The UK.