Terminated

Abbott Point of Care i-Stat cTnI cartridge. The i-Stat cardiac troponin I (cTnI) test is an in vitro diagnostic test for the quantitive measurement of cardiac troponin I (cTnI) in whole blood or plasma. Measurements of cardiac troponin I are used in the diagnosis and treatment of myocardial infarction and as an aid in the risk stratification of patients with acute coronary syndromes with respect to the relative risk of mortality.

FDA Recall Z-0716-2013 · Posted 2013-01-24

Recall StatusTerminated
Product CodeMMI
Date Initiated2012-11-01
Date Terminated2013-06-14
Product Quantity560 boxes (25 cartridges in each box)

Reason for Recall

APOC has determined that 10 boxes of i-Stat cTnI cartridges from lot P12179 have an incorrect barcode applied to the cartridge portion pack.

Firm Action

APOC sent an "IMPORTANT PRODUCT INFORMATION" letter dated October 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Abbott Point Of Care Technical Support at 800-366-8020, Option 1 or your Abbott Point Of Care representative for questions regarding this information.

Root Cause

Error in labeling

Distribution Pattern

Worldwide Distribution.